Error: 30-01 (Partial fill error)

Troubleshooting partial fill error 30-01

Possible cause one: Not enough sample has been applied to the test strip porch.  The sample for testing should completely fill the test strip's semi-circular white notch.

Possible cause two:  The sample is spritzy and requires a degas step first.  CO₂ bubbles can displace sample on the strip.

Possible cause three: The strips in use have been compromised. Please trial a different lot #.

Note: If you receive a 30-01 error, do NOT apply additional sample to the test strip after the test has begun.  This will not solve the issue.

If the error persists, contact your Supplier with a detailed description of the problem.  Include Device Serial Number and Software Number details.