Error: 70-21 (Sample not applied)

Troubleshooting error 70-21

Possible cause 1: The sample was not applied during the time requested by the analyzer. 

Discard the strip before starting a new test. Apply the sample when prompted by the analyzer.

Possible cause 2: The sample contains excessive carbon dioxide and needs to be degassed.

Degas sample, then retest.

Possible cause 3: Wine previously tested has seeped into the strip port, affecting the meter's ability to detect sample. Tip: See section 4, “Cleaning the analyzer” in the Sentia User Guide.

  1. Remove the protective cap that sits over the Sentia strip port.
  2. Using absorbent, lint free paper, carefully blot up the excess liquid from the inside of the protective cap. Note: While cleaning spills, always keep the strip port pointed downwards to avoid getting liquid into the analyzer.
  3. Using a new strip of lint free paper each time, continue to insert the paper into the strip port until it comes out clean.
  4. If the error still persists, let the device dry overnight.

Possible cause 4: Meter contact pins dirty.

Insert and remove a used strip to clean contactors of the meter. Then try a test again.